Skip to content

canonical: https://aviantestpro.com/knowledge/quality-control/accuracy-validation

Method Validation & Accuracy Assurance

All testing methods at SENO undergo a standardized validation process before clinical use. The validation framework follows Clinical and Laboratory Standards Institute (CLSI) EP guidelines and meets the requirements of our ISO 9001:2015 and ISO 13485:2016 certified quality management systems.


Validation Framework

SENO's validation workflow consists of three phases, each documented in a validation file that is reviewed and approved by both R&D and Quality Management.

PhaseActivitiesDeliverableTimeline
Phase 1 — Analytical ValidationLOD, specificity, precision (repeatability + reproducibility), linearity, amplification efficiency, LOQValidation Plan + Analytical Data Report4–8 weeks
Phase 2 — Diagnostic ValidationClinical sample panel testing (minimum 100 clinical samples), sensitivity/specificity calculation, ROC analysisDiagnostic Accuracy Report4–12 weeks
Phase 3 — Ongoing MonitoringInternal QC, proficiency testing, retest rate tracking, complaint investigation, QC chart monitoringQuarterly QC ReviewPerpetual

Validation Parameters

The table below defines the core validation parameters measured for every quantitative assay (qPCR). Conventional PCR assays (e.g., CHD sexing) are validated on concordance rather than quantitative metrics.

ParameterAbbreviationDefinitionSENO Acceptance CriterionExample (PBFD qPCR)
Analytical Sensitivity (Limit of Detection)LODLowest concentration detected with ≥ 95% confidenceDocumented per assay10 copies/reaction
Analytical SpecificityNo cross-reactivity with related or common co-infecting organismsPanel of 6+ organisms tested9 organisms tested, all negative
Precision (Repeatability)rSame-day, same-operator, same-instrument variationCV < 3% (Ct)0.6% CV
Reproducibility (Inter-assay)RDay-to-day, operator-to-operator variationCV < 5% (Ct)0.96% CV
Diagnostic SensitivityDSeProportion of true positives correctly identified≥ 95%95.9%
Diagnostic SpecificityDSpProportion of true negatives correctly identified≥ 95%98.0%
Area Under ROC CurveROC-AUCOverall ability to discriminate positive from negative≥ 0.950.97
Linearity (Dynamic Range)Range over which target concentration is proportional to signalR² ≥ 0.98R² = 0.998
Amplification EfficiencyEPCR efficiency calculated from standard curve slope90–110%100.1%
Limit of QuantificationLOQLowest concentration quantifiable with CV < 3% (Ct)CV < 3% (Ct)100 copies
Inter-laboratory ReproducibilityResults consistent across different laboratoriesQualitative: 100% agreement; quantitative: CV < 10%In progress

Parameter Measurement Methods

ParameterCLSI ReferenceMethod
LODEP-17 A2Probit analysis from 60+ replicates across 5+ concentrations near estimated detection limit
PrecisionEP-15 A320 replicates of 3 concentrations over 5 days
LinearityEP-6 A28–10 concentration points across dynamic range; linear regression
Diagnostic accuracyEP-12 A2 (qualitative), EP-24 A3 (quantitative)Clinical panel with reference standard; McNemar's test for paired comparison

Validation by Test Type

Each test category follows a tailored validation approach reflecting the assay format and intended use.

Test TypeAssay FormatValidation ApproachKey MetricsStatus
DNA Sexing (CHD PCR)Conventional PCR + GelKnown-sex birds (50+ species, 200+ individuals); Sanger sequencing of ambiguous results≥ 99% concordanceValidated
PBFD / BFDV DetectionqPCR (TaqMan)CLSI EP-17 (LOD), EP-15 (precision), EP-12 (qualitative); 138 clinical samplesLOD 10 copies; DSe 95.9%; DSp 98.0%Validated
APV DetectionqPCR (TaqMan)Same CLSI framework; spiked tissue samples + clinical panelLOD 20 copies; DSe ≥ 95%Validated
AIV DetectionqPCR (TaqMan)OIE-referenced primer/probe set; avian influenza subtypes panelLOD 15 copies; DSe ≥ 95%Validated
DNA FingerprintingMultiplex STR CapillaryPopulation allele frequency in 15+ populations; Mendelian inheritance checkDP > 0.9999; PI > 10⁶Validated
Performance Gene Testing (Pigeons)SNP GenotypingPopulation association study (2,500 racing pigeons); independent replication (800 birds)Significant p < 0.01Validated
Pigeon Disease Panel (11-plex)Multiplex PCR + RT-PCRTarget-specific LOD for each of 11 pathogens; 200+ clinical samples panelPer-target LOD 10–50 copiesValidated

Ongoing Accuracy Monitoring

Even after initial validation, every assay is continuously monitored through a multi-layered QC system.

MonitorFrequencyMethodTargetResponsible
Internal batch QCEvery batchPositive / negative / NTC / IAC / extraction blank100% passLaboratory Technician
Intra-laboratory duplicate5% of samplesBlind duplicate testing by separate analyst100% agreementQuality Manager
Inter-laboratory proficiencyAnnuallyExternal accredited PT provider (e.g., RCPA, QCMD)100% passQuality Manager
Retest rateMonthlyNumber of retests / total samples × 100%< 2%Quality Manager
Client discrepancy rateContinuousComplaints received / total reports × 100%< 0.5%Quality Manager
QC chart monitoring (Levey-Jennings)Continuous per control lotPlot Ct values against established mean ± 3 SDNo Westgard rule violationsLaboratory Supervisor
Method comparison (new vs. old)On method change40+ paired clinical samples, Bland-Altman analysisBias ≤ 5% of meanR&D Team

Levey-Jennings QC Rules Applied

Westgard RuleDescriptionAction
1₂sOne control value exceeds ± 2 SDWarning; inspect controls
1₃sOne control value exceeds ± 3 SDReject batch; investigate root cause
2₂sTwo consecutive values exceed ± 2 SD (same side)Reject batch; systematic error
R₄sTwo controls differ by > 4 SDReject batch; random error
10xTen consecutive values on same side of meanReject batch; trend — calibrate instrument

Verification vs. Validation

It is important to distinguish between verification and validation, as both are required under ISO 13485.

TermDefinitionSENO ApplicationRegulatory Reference
VerificationConfirming that specified requirements have been fulfilled through objective evidenceEach production lot of IVD reagents is functionally tested against a reference lot before releaseISO 13485 § 7.3.6
ValidationConfirming that the method is suitable for its intended use through objective evidenceComplete CLSI EP validation protocol performed before clinical deployment of any new assayISO 13485 § 7.3.7

When Re-Validation Is Required

A full or partial re-validation is triggered under the following circumstances:

ChangeRe-Validation ScopeMinimum Sample Size
New primer/probe set for existing targetFull analytical validation (LOD, specificity, precision, linearity)60+ replicates
New instrument platform (e.g., different qPCR model)Precision + linearity comparison (40+ samples)40 paired samples
New reagent supplier or formulationEquivalency study (40+ paired samples)40 paired samples
SOP revision affecting critical parametersPartial re-validation of affected parametersPer parameter
PT failure or quality incidentInvestigation-led — may trigger full re-validationDetermined by investigation
More than 12 months since last validationPrecision re-check (repeatability + reproducibility)20 replicates
Change in sample matrix (e.g., feather to blood)Matrix equivalency study20 paired samples per matrix

Laboratory Accreditation and External Audits

Accreditation / AuditBodyScopeFrequencyLast Completed
ISO 9001:2015TÜV Rheinland / SGSQuality management systemAnnual surveillance + 3-year recertification2026 Q1
ISO 13485:2016TÜV Rheinland / SGSMedical device QMSAnnual surveillance + 3-year recertification2026 Q1
CMA (China Metrology Accreditation)CNASLaboratory testing competenceAnnual2025 Q4
CNAS-CL02 (ISO 15189 equivalent)CNASMedical laboratory qualityBiennial2026 Q2
Proficiency testing programmesRCPA / QCMD / external PT providerInter-laboratory comparisonAnnually per assay2026 Q2

Cross-References


This document is maintained by SENO's R&D and Quality Management teams. Validation files are controlled documents. For enquiries, contact quality@senobio.com.